Multi-sensor coordinate unification in dimensional metrology is used in order to get holistic, more accurate and reliable information about a workpiece based on several or multiple measurement values from one or more sensors. Because of the problem that standard ball is deficient as a standard artifact in the coordinate unification of high-precision composite measurement in two dimensions (2D) , a new method is proposed in this paper which uses angle gauge blocks as standard artifacts to achieve coordinate unification between the image sensor and the tactile probe. By comparing the standard ball with the angle gauge block as a standard artifact, theoretical analysis and experimental results are given to prove that it is more precise and more convenient to use angle gauge blocks as standard artifacts to achieve coordinate unification of high-precision composite measurement in two dimensions.