Objective: To evaluate the therapeutic effect, safety and complications oftranscathetering Amplatzer device in the closure of patent ductus arteriosus (PDA) and atrial septaldefects (ASD) in children. Methods: Patients with PDA (n = 25) and ASD ( n = 16), confirmed byechocardiography, were treated by transcatheterization. Amplatzer occluder device was placed by thetranscatheterization with the image support of X-ray and transthoracic echocardiography ( TTE) . TheTTE, ECG and X-ray examination were engaged to evaluate the therapeutic results on the time pointsof 24 h , 1, 3 , 6, 12 months after the operation, and all these cases were engaged to the follow-upexamination. Results: The cardioangiographic diameter was 13.0-28.0 mm ([19.3+-4.9] mm) in ASD and2.0-7.7 mm ([3.9+-1.5] mm) in PDA . The diameter of the Amplatzer occluder selected were 13.0-30.0mm ( [20.6+-5.1] mm) in ASD and 4.0-12.0 mm ([6.6+-1.9] mm) in PDA , respectively. All the Amplatzeroccluders were placed successfully. There were no complications during and after the operation.Very small residual shunt was still found soon after the operation in 10 cases , and there were noresidual or recanalization after three months of the operation. The pulmonary artery pressure andheart size were significantly decreased in follow-up examination . Conclusion: Transcatheteringclosure with Amplatzer device is an effective, simple and safe technique in the treatment of ASD andPDA in children.